Before You Start
Safety Information
1.3 Safety Information
Use this instrument only as specied in this manual. Otherwise, the protection
provided by the instrument may be impaired. Refer to the safety information in the
Warnings and Cautions sections below.
The following denitions apply to the terms “Warning” and “Caution”.
“Warning” identies conditions and actions that may pose hazards to the user.
“Caution” identies conditions and actions that may damage the instrument
being used.
1.3.1 Warnings
To avoid personal injury, follow these guidelines.
GENERAL
DO NOT use this instrument in environments other than those listed in the
Users Guide.
Inspect the instrument for damage before each use. DO NOT use the instrument
if it appears damaged or operates abnormally.
Follow all safety guidelines listed in the users manual.
Calibration Equipment should only be used by Trained Personnel.
If this equipment is used in a manner not specied by the manufacturer, the
protection provided by the equipment may be impaired.
Before initial use, or after transport, or after storage in humid or semi-humid
environments, or anytime the dry-well has not been energized for more than
10 days, the instrument needs to be energized for a “dry-out” period of 2 hours
before it can be assumed to meet all of the safety requirements of the IEC
1010-1. If the product is wet or has been in a wet environment, take necessary
measures to remove moisture prior to applying power such as storage in a low
humidity temperature chamber operating at 50 degree centigrade for 4 hours or
more.
DO NOT use this instrument for any application other than calibration work.
The instrument was designed for temperature calibration. Any other use of the
instrument may cause unknown hazards to the user.
Completely unattended operation is not recommended.
Overhead clearance is required. DO NOT place the instrument under a cabinet
or other structure. Always leave enough clearance to allow for safe and easy
insertion and removal of probes.
If the instrument is used in a manner not in accordance with the equipment
design, the operation of the dry-well may be impaired or safety hazards may
arise.
This instrument is intended for indoor use only.