9100S Dry-Well Calibrator
Serial Interface Parameters
26
bAUd FlashesbAUdforonesecondandthendisplaysthe
setting
2400b Currentbaudrate
The baud rate of the serial communications may be programmed to 300, 600, 1200,
2400 (default), 4800, or 9600 baud. Use the up or down arrows to change the baud rate
value.
4800 b Newbaudrate
Press “SET” to accept the new setting or “EXIT” to abort the operation and skip to the
next parameter in the menu.
6.11.2 Sample Period
The sample period is the next parameter in the serial interface parameter menu. The
sample period is the time period in seconds between temperature measurements
transmitted from the serial interface. If the sample rate is set to 5, the instrument
transmits the current measurement over the serial interface approximately every ve
seconds. The automatic sampling is disabled with a sample period of 0. The sample
period is indicated by,
SPEr Flashesforonesecondandthentheserialsampleperiod
settingisdisplayed
1 Currentsampleperiod(seconds)
Adjust the value by using the up or down arrows (u d).
60 Newsampleperiod
Press “SET” to accept the new setting or “EXIT” to abort the operation and skip to the
next parameter in the menu.
6.11.3 Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full duplex or
half duplex. With full duplex any commands received by the calibrator via the serial
interface are immediately echoed or transmitted back to the device of origin. With half
duplex the commands are executed but not echoed. The duplex mode parameter is
indicated by,
dUPL Flashesforonesecondandthentheserialduplexmode
settingisdisplayed
FULL Currentduplexmodesetting
The mode may be changed using the up or down arrows (u d).