Chapter 2 To Set the Vertical System RIGOL
DS1000Z User Guide 2-7
Delay Calibration of the Analog Channel
When using an oscilloscope for actual measurement, the transmission delay of the
probe cable may bring relatively greater error (zero offset). DS1000Z supports user
to set a delay time to calibrate the zero offset of the corresponding channel. Zero
offset is defined as the offset of the crossing point of the waveform and trigger level
line relative to the trigger position, as shown in the figure below.
Figure 2-1 Zero Offset
Press CH1  Delay-Cal and use to set the desired delay time. The range
available is from -100 ns to 100 ns. Pressing down the multifunction knob will
reset the delay time to 0.00 s.
Note: This parameter is related to the instrument model and the current horizontal
time base setting. The larger the horizontal time base is, the larger the setting step
will be. Take DS1104Z-S Plus as an example. The step values under different
horizontal time base are as shown in the table below.
Table 2-2 Relation between delay calibration time step and horizontal time base
Horizontal Time Base
Delay Calibration Time Step
5 ns
100 ps
10 ns
200 ps
20 ns
400 ps
50 ns
1 ns
100 ns
2 ns
200 ns
4 ns
500 ns
10 ns
1 μs to 10 μs
20 ns
Note: When the horizontal time base is equal to or greater than 10 μs, the delay
calibration time cannot be adjusted.