Chapter 13 Signal Source RIGOL
DS1000Z User Guide 13-15
7. Delete
Press Delete to delete the current point from the waveform and connect the
remaining points using the current interpolation mode.
Note: Point 1 cannot be deleted.
8. Apply
Press Apply to finish editing the current waveform.
9. Save
Press Save to enter the file store interface. Please refer to the introduction in
"
Store and Recall" to save the current waveform file in ".arb" format in the
internal memory or external memory (you can overwrite the original file or save
the current waveform again).You can select the arbitrary waveforms in the
internal or external memory for output. For the details, refer to the introduction
in "To Select Waveform".
Modulation
The built-in signal generator of DS1000Z series oscilloscope supports amplitude
modulation (AM) and frequency modulation (FM). The modulated waveform consists
of the carrier waveform and modulating waveform. The carrier waveform is the
waveform output from the signal generator and the modulating signal can be the
built-in sine, square, ramp waveform or noise signal.
Press Src1Mod to open the source 1 modulation setting menu. Press Modulation
to turn on or off the modulation function. Press Mod.Type to set the modulation
type of the current signal to "AM" or "FM". Then, you can set the modulation
parameters according to the selected modulation type.