Chapter 1 Quick Start RIGOL
DS1000Z User Guide 1-7
Function Inspection
1. Press Storage  Default to restore the oscilloscope to its default
configuration.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in the figure below.
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe.
Figure 1-7 To Use the Compensation Signal
4. Set the attenuation on the probe to 10X. Then press AUTO.
5. Observe the waveform on the display. In normal condition, the square waveform
as shown in the figure below should be displayed.
Figure 1-8 Square Waveform
6. Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform "Probe
Compensation".
WARNING
To avoid electric shock when using the probe, please make sure that the
insulated wire of the probe is in good condition. Do not touch the metallic
part of the probe when the probe is connected to high voltage source.
Tip
The signal output from the probe compensation connector can only be used for
probe compensation adjustment and cannot be used for calibration.
Compensation Signal Output Terminal
Ground Terminal