FLUKE 190-204/AM/S ScopeMeter Test Tool with SCC-290 kit, 200 MHz, 4 channels

User Manual for the Fluke 190-204/AM/S ScopeMeter Test Tool with SCC-290 kit, 200 MHz, 4 channels - Page 36

For 190-204/AM/S.

PDF File Manual, 150 pages, Read Online | Download pdf file

190-204/AM/S photo
Loading ...
Loading ...
Loading ...
Using the Scope and Meter
Acquiring Waveforms
1
29
3
ENTER
Jump to Waveform: and select
Mathematics... to open the
Mathematics menu.
4
ENTER
Select Function: Spectrum.
5
ENTER
Select the source waveform for the
spectrum: Source : A, B, C or D
6
ENTER
Select Window: Auto (automatic
windowing), Hanning, Hamming,
or None (no windowing).
You will see a screen that looks like Figure 11.
Observe that the top right of the screen displays
SPECTRUM.
If it displays LOW AMPL a spectrum measurement cannot
be done as the waveform amplitude is too low.
If it displays WRONG TB the time base setting does not
enable the test tool to display an FFT result. It is either too
slow, which can result in aliasing, or too fast, which results
in less than one signal period on the screen.
7
F1
Perform a spectrum analysis on
trace A, B, C or D.
8
F2
Set the horizontal amplitude scale
to linear or logarithmic.
9
F3
Set the vertical amplitude scale to
linear or logarithmic.
10
F4
Turn the spectrum function off/on
(toggle function).