FLUKE 190-204/AM/S ScopeMeter Test Tool with SCC-290 kit, 200 MHz, 4 channels

User Manual for the Fluke 190-204/AM/S ScopeMeter Test Tool with SCC-290 kit, 200 MHz, 4 channels - Page 116

For 190-204/AM/S.

PDF File Manual, 150 pages, Read Online | Download pdf file

190-204/AM/S photo
Loading ...
Loading ...
Loading ...
Specifications
Oscilloscope
8
111
5 ns to 4 μs /div (all channels)............ up to 1.25 GS/s
10 μs to 120 s/div .........................................125 MS/s
FLUKE 190-062:
10 ns to 4 μs /div (all channels)...........up to 625 MS/s
10 μs to 120 s/div ........................................ 125 MS/s
Record Length: see table below.
Table 2. Record Lenth (Samples/points per input)
Mode Glitch Detect On Glitch Detect Off Max. Samplerate
Scope - Normal 300 min/max pairs 3k true samples compressed
into 1 screen (300 samples per
screen)
Scope - Fast 300 min/max pairs 300 true samples
Scope - Full 300 min/max pairs 10k true samples, compressed
into 1 screen.
Use Zoom and Scroll to see
waveform details
190-062: 625 MS/s
190-102/104: 1.25 GS/s
190-202/204: 2.5 GS/s (1 or 2
channels on)
190-204: 1.25 GS/s (3 or 4
channels on)
190-502: 5 GS/s (1 channel on)
190-502: 2.5 GS/s (2 channels
on)
Scope Record Roll 30k samples 4x 125 MS/s
Trend Plot > 18k min/max/ average values
per measurement
Up to 5 measurements per second
Glitch Detection
5 μs to 120 s/div........... displays glitches as fast as 8 ns
Waveform Display ............................................A, B, C, D,
Math (+, -, x, X-Y mode, spectrum)
Normal, Average, Persistence, Reference
Time Base Accuracy ......................±(100 ppm + 0.04 div)