FLUKE 190-204/AM/S ScopeMeter Test Tool with SCC-290 kit, 200 MHz, 4 channels

User Manual for the Fluke 190-204/AM/S ScopeMeter Test Tool with SCC-290 kit, 200 MHz, 4 channels - Page 26

For 190-204/AM/S.

PDF File Manual, 150 pages, Read Online | Download pdf file

190-204/AM/S photo
Loading ...
Loading ...
Loading ...
Using the Scope and Meter
Making Automatic Scope Measurements
1
19
Making Automatic Scope Measurements
The test tool offers a wide range of automatic scope
measurements. In addition to the waveforms you can
display four numeric readings: R
EADING 1 4. These
readings are selectable independently, and the
measurements can be done on the input A , input B, input
C or input D waveform.
To choose a frequency measurement for input A, do the
following:
1
SCOPE
Display the SCOPE key labels.
2
F2
Open the READING .. menu.
3
F1
Select the reading number to be
displayed, for example READING 1
4
ENTER
Select on A. Observe that the
highlight jumps to the present
measurement.
5
ENTER
Select the Hz measurement.
Observe that the top left of the screen displays the Hz
measurement. (See Figure 8.)
To choose also a Peak-Peak measurement for Input B as
second reading, do the following:
1
SCOPE
Display the SCOPE key labels.
2
F2
Open the READING .. menu.
3
F1
Select the reading number to be
displayed, for example READING 2
4
ENTER
Select on B. The highlight jumps
to the measurements field.
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct
.com
Loading ...
Loading ...
Loading ...