FLUKE 190-204/AM/S ScopeMeter Test Tool with SCC-290 kit, 200 MHz, 4 channels

User Manual for the Fluke 190-204/AM/S ScopeMeter Test Tool with SCC-290 kit, 200 MHz, 4 channels - Page 133

For 190-204/AM/S.

PDF File Manual, 150 pages, Read Online | Download pdf file

190-204/AM/S photo
Loading ...
Loading ...
Loading ...
Fluke 190 Series II
Users Manual
128
Electromagnetic Immunity
The Fluke 190 Series II test tools, including standard accessories, conform with the EEC directive 2004/108/EC for EMC
immunity, as defined by EN-61326-1, with the addition of the following tables.
Trace disturbance with VPS410/VPS510 voltage probe shorted (Scope Mode, 10 ms/div):
Table 3. (E = 3V/m)
Frequency No Disturbance Disturbance < 10% of full
scale
Disturbance > 10% of full
scale
80 MHz – 700 MHz All other ranges 100, 200, 500 mV/div 2, 5, 10, 20, 50 mV/div
700 MHz – 1 GHz All other ranges 10 mV/div 2, 5 mV/div
1.4 GHz – 2.7 GHz All ranges
Meter Mode (Vdc, Vac, Vac+dc, Ohm and Continuity): Reading disturbance with test leads shorted
Table 4
No visible disturbance E = 3V/m
Frequency range 10 kHz to 1 GHz 500 mV to 1000 V , 500 Ohm to 30 MOhm ranges
Table 5
No visible disturbance E = 3V/m
Frequency range 1.4 GHz to 2 GHz 500 mV to 1000 V, 500 Ohm to 30 MOhm ranges