SDS6000A Series Digital Oscilloscope User Manual
int.siglent.com 239
Full Screen: Time-domain waveform and frequency-domain waveform are
displayed together.
Exclusive: Only the frequency-domain waveform is displayed.
Split Mode, Zoom off
Split Mode, Zoom on
Full-Screen Mode
Exclusive Mode
FFT Mode
Normal: Displays the FFT result of each frame directly.
Max-Hold: Holds the maximum value in the historic frame on the display
until cleared. This mode is suitable for detecting discontinuous waves, such
as sporadic pulse signals, or frequency hopping signals. Press the Clear
Sweeps button on the front panel to clear the max-hold waveform.
Average: Reduces the effect of the superimposed random noise on the
signal. After the FFT mode is set as Average, Average will appear under
FFT Mode. The average count can be set here, ranging from 4 to 1024.
Press the Clear Sweeps on the front panel or touch Reset in the dialog box