On the system settings interface users can set the language, auto power off, brightness, power
setting and buzzer. And view the model and version of the instrument.
Press [SET] key to enter the system setting interface, press the arrow key to select the setting
option and highlight, and press the arrow key for specific setting. Then press the [SET] button
again to view the model and version of the instrument. Press [SET] key again to exit the system
setting interface.
Power on setting:
Default: If the power on setting is <default>, all setting will be system default for next start.
Last: If the boot setting is <Last>, the test parameters, system language, automatic shutdown,
back light setting and beep setting will be saved for the next start.
1.3 Test port
The serial LCR uses the 3-terminal and 5-terminal test ports at the same time, which is to
combine convenience and high accuracy for the test. See the figure for the test terminal.
Complete
five terminal
test hole
Access three
terminal test
hole
HI
LO
GUARD
The three-terminal test port of the instrument uses the standard rubber jack, therefore the
inexpensive rubber plug- alligator clip can be used as the test line. It is very convenient to apply
the extended test, but the drawback is low testing accuracy.
To improve the accuracy of the test line when using the extension line, the LCR is also equipped
with the five-terminal test notch for dedicated test fixture. It renders possible the complete
four-terminal measurement of the extension line to ensure the high testing accuracy.