Loading ...
Loading ...
Loading ...
![](https://files.manualsfile.com/16523235-hantek1832c-hantek1833c/bg1d.png)
27
Impedance Z and phase angle
◼ 100Hz/120Hz/400Hz/1kHz/4KHz/10kHz
Range
Range of display
Accuracy Ze
Accuracy
e
Equivalent mode
recommended
20MΩ
4.000MΩ~20.000MΩ
3.0%+10 digits
3.4°
Parallel
4MΩ
400.0kΩ~3.9999MΩ
1.2%+3 digits
0.7°
Parallel
400kΩ
40.00kΩ~399.99kΩ
0.3%+3 digits
0.2°
Parallel
40kΩ
4.000kΩ~39.999kΩ
0.25%+2 digits
0.1°
------
4kΩ
400.0Ω~3.9999kΩ
0.25%+2 digits
0.1°
Series
400Ω
40.00Ω~399.99Ω
0.25%+2 digits
0.1°
Series
40Ω
4.000Ω~39.999Ω
0.5%+3 digits
0.3°
Series
4Ω
0.4000Ω~3.9999Ω
2.0%+3 digits
1.1°
Series
0.4Ω
0.0000Ω~0.3999Ω
4.0%+3 digits
------
Series
◼ 40kHz/50KHz
Range
Range of display
Accuracy Ze
Accuracy
e
Equivalent mode
recommended
20MΩ
4.000MΩ~20.000MΩ
7.0%+41 digits
4.0°
Parallel
4MΩ
400.0kΩ~3.9999MΩ
2.5%+3 digits
1.4°
Parallel
400kΩ
40.00kΩ~399.99kΩ
1.0%+4 digits
0.6°
Parallel
40kΩ
4.000kΩ~39.999kΩ
1.0%+4 digits
0.6°
------
4kΩ
400.0Ω~3.9999kΩ
0.5%+3 digits
0.3°
Series
400Ω
40.00Ω~399.99Ω
0.5%+3 digits
0.3°
Series
40Ω
4.000Ω~39.999Ω
0.7%+4 digits
0.4°
Series
4Ω
0.4000Ω~3.9999Ω
2.0%+6 digits
1.1°
Series
0.4Ω
0.0000Ω~0.3999Ω
5.0%+10 digits
------
Series
◼ 75KHz/100kHz
Range
Range of display
Accuracy Ze
Accuracy
e
Equivalent mode
recommended
20MΩ
4.000MΩ~20.000MΩ
9.0%+20 digits
5.2°
Parallel
4MΩ
400.0kΩ~3.9999MΩ
4.0%+10 digits
2.3°
Parallel
400kΩ
40.00kΩ~399.99kΩ
1.5%+4 digits
0.9°
Parallel
40kΩ
4.000kΩ~39.999kΩ
1.0%+2 digits
0.6°
Parallel
4kΩ
400.0Ω~3.9999kΩ
0.7%+2 digits
0.4°
------
400Ω
40.00Ω~399.99Ω
0.7%+2 digits
0.4°
Series
40Ω
4.000Ω~39.999Ω
1.0%+5 digits
0.6°
Series
4Ω
0.4000Ω~3.9999Ω
3.0%+10 digits
1.7°
Series
0.4Ω
0.0000Ω~0.3999Ω
7%+20 digits
------
Series
Loading ...
Loading ...