Probe Compensation
When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement inaccuracy or error. The probe
compensation procedures are as follows.
1. Perform steps 1, 2, 3 and 4 of Function Inspection in the previous section.
2. Check the waveforms displayed and compare them with the following:
Under Perfectly Over
Compensated Compensated Compensated
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the waveform displayed is as the Perfectly
compensated” in the figure above.