SDS2000X-E Series User Manual - Page 220

For SDS2000X-E.

Loading ...
Loading ...
Loading ...
on the scope , set CH2 to DC coupling, open bandwidth limit, and probe
attenuation to 1X.
b. Adjust the voltage of channel 2 channel to 1 mV/div, and make the
waveform visible in the oscilloscope interface, press the Measure and
measure the mean value of channel 2.
c. Press the Manual and rotate the knob to adjust the waveform, so that the
mean value of the observed CH2 will be in the range of ± 1 mV, and then
press the Save. Upon completion of the zero adjust process, the
oscilloscope will display " Zero Adjust Completed! ".
Systems
1. System Status
Press Utility NextPage4/4 AWG System , then select the System Status
option to view the AWG’s system information, including software version, hardware
version, and serial number. After finished, press the return soft key to exit the
system information page.
Figure 115 AWG System Information Interface
2. Update
The software version file of the AWG can be updated directly via U-disk. Follow the
Loading ...
Loading ...
Loading ...