Note: The property cannot be set when the list is "Run". The Run/Stop trigger output is valid
only when the trigger source is set to Key. When using the list output, you should turn on
the channel output first, and then press "Run/Stop", otherwise the complete list cannot be
output.
8. Data logger
The data recording function is used to view and record the output voltage, current, and
power waveforms and data. This function must be used with the external USB storage
device on the front panel, the waveform is displayed on the screen, and the data is directly
recorded to the external storage device.
Data record setting steps:
Press the "Data Logger" button in the menu bar to enter the data logger setting interface.
1. Select the data output channel and set the output parameters.
2. Set the waveform parameters
All recordable waveform parameters are displayed on the left side of the screen. Press the
arrow keys to select the parameter to be recorded, and press Enter to confirm. Press
Waveform Settings to set the waveform vertical parameters and horizontal parameters
respectively.
Vertical: Set the voltage/current/power value represented by the unit grid in the vertical
direction of the displayed waveform, the setting range is 10uV50V/10uA50A/10uW
50W, press "Volts/Div +", "Volts/Div -" or " Amps/Div +", "Amps/Div -" or "Watts/Div +",
"Watts/Div -" keys to adjust the size, the step is 1-2-5.
Press "Offset +", "Offset -" to set the vertical offset of the displayed waveform.