4. Output coupling
Press Source Settings->On/Off Coupling to set the output coupling. After setting the
coupling, turn on the output of any one of the channels, and the other coupling channels will
be turned on at the same time. If only part of the channel output coupling is set, open the
front panel [All On/Off], regardless of whether the channel is set to output coupling, all
channels will be turned on.
Note: Single channel output, no need to set output coupling.
5. Output inhibit
Press Output Settings->Output Inhibit to set the inhibition mode (off, latched or live) and
DIO Pin3 function.
Inhibit Off: Inhibit input is ignored.
Inhibit Latched: Inhibit the transition from True to False in the input state of the external
logic signal, and always inhibit the output.
Inhibit Live: Allow the transition of the external logic signal input state from True to False.
The output state is determined by the external input signal. When the external input signal
is True, the output is disabled; when the external input signal is False, the output is
enabled.
DIO Pin3: Can be set to remote inhibit input or digital input.
INH: Remote inhibit input, the output status of all channels of the power supply can be
controlled through external signals.
DINP: Digital input, you can trigger the input by setting the level of Pin3.