Binder VD 56 vacuum drying chamber for non-flammable solvents

User Manual - Page 81

For VD 56. Also, The document are for others Binder models: VD (E3.1)

PDF File Manual, 179 pages, Read Online | Download pdf file

VD 56 photo
Loading ...
Loading ...
Loading ...
VD (E3.1) 10/2020 page 81/179
10.1.6 Setting the chamber address
This setting is required for the communication with the BINDER APT-COM™ 4 Multi Management
Software. The chamber address settings in the chamber controller and in the software must be identical.
Required access level: “Admin”.
Path: Normal display Settings Chamber Chamber address
Press the OK button to enable the setting.
Setting the chamber address
The current setting flashes. Enter the desired address with the arrow
buttons. Setting range: 1 up to 254
Confirm the entry with the OK button.
Chamber address
With the arrow-down button you can now change to the next parameter (display brightness).
With the Back button you can go back to the “Chamber” submenu and, repeatedly pressing it, to
Normal display.
10.1.7 Display brightness
Required access level: “Admin”.
Required access level: “Admin”.
Path: Normal display Settings Chamber Brightness
Press the OK button to enable the setting.
Setting the display brightness
The current setting flashes. Enter the desired value with the arrow
buttons. Setting range: 10% up to 100%
Confirm the entry with the OK button.
Brightness
With the arrow-down button you can now change to the next parameter (audible alarm, chap. 13.4.1).
With the Back button you can go back to the “Chamber” submenu and, repeatedly pressing it, to
Normal display.
10.1.8 Display of the inner chamber temperature and the heating temperature
Path: Normal display keep pressed down for 5 seconds.
The current inner chamber temperature (upper value) and the temperature measured directly on the
heating element (lower value) are displayed.
T / pressure (mbar)
Interior / heating
Press down the Back button again for 5 seconds to terminate the indication.