Binder VD 56 vacuum drying chamber for non-flammable solvents

User Manual - Page 151

For VD 56. Also, The document are for others Binder models: VD (E3.1)

PDF File Manual, 179 pages, Read Online | Download pdf file

VD 56 photo
Loading ...
Loading ...
Loading ...
VD (E3.1) 10/2020 page 151/179
25. Maintenance and service, troubleshooting, repair, testing
25.1 General information, personnel qualification
Maintenance
See chap. 25.3
Simple troubleshooting
Chap. 25.2 describes troubleshooting by operating personnel. It does not require technical intervention
into the chamber, nor disassembly of chamber parts.
For personnel requirements please refer to chap. 1.1.
Detailed troubleshooting
If errors cannot be identified with simple troubleshooting, further troubleshooting must be performed by
BINDER Service or by BINDER qualified service partners or technicians, in accordance with the
description in the Service Manual.
For personnel requirements please refer to the Service Manual.
Repair
Repair of the chamber can be performed by BINDER Service or by BINDER qualified service partners
or technicians, in accordance with the description in the Service Manual.
After maintenance, the chamber must be tested prior to resuming operation.
Electrical testing
To prevent the risk of electrical shock from the electrical equipment of the chamber, an annual repeat
inspection as well as a test prior to initial startup and prior to resuming operation after maintenance or
repair, are required. This test must meet the requirements of the competent public authorities. We
recommend testing under DIN VDE 0701-0702:2008 in accordance with the details in the Service
Manual.
For personnel requirements please refer to the Service Manual.
25.2 Simple troubleshooting
Defects and shortcomings can compromise the operational safety of the chamber and can lead to risks
and damage to equipment and persons. If there are is a technical fault or shortcoming, take the chamber
out of operation and inform BINDER Service. If you are not sure whether there is a technical fault,
proceed according to the following list. If you cannot clearly identify an error or there is a technical fault,
please contact BINDER Service.
Only qualified service personnel authorized by BINDER must perform repair.
Repaired chambers must comply with the BINDER quality standards.