8. Setting special controller functions
You can set the following functions via the controller menu:
Activate / deactivate the Standby mode (chap. 6.2)
Turning off the vacuum pump via the socket on the pump module (chap. 8.2)
Use the optional connectionGAS/AIR2” (5) for ventilation (chap. 8.3)
Close all existing fine dosing valves (chap. 8.4)
Turning on / off temperature control (chap. 8.5)
Turning on / off pressure control (chap. 8.6)
Starting / terminating the drying monitoring (chap. 8.7)
8.1 Menu structure
8.1.1 RD4 controller
Required access level: “User”.
8.1.1.1 Functions on/off” menu
In the “Functions on/off” menu you can define the switching state of three special controller functions.
Path: Normal display Setpoints Functions on/off
The functions are displayed from left to right.
Example: Function 1 activated = 1000. Function 1 deactivated = 0000.
Submenu “Functions on/off”.
This view shows the switching states four functions.
Functions 1, 2,
and 3 are available with the present chamber type.
“1” = Function activated
“0” = Function deactivated
Functions on/off
Press the OK button to access the first individual function. With the arrow-down button you can
proceed to the subsequent functions.
Function 1 “Standby” (chap. 6.2)
Function 2 GAS/AIR 2” (chap. 8.3)
Function 3 Close all valves (chap. 8.4)
Function 4 “Vacuum pump off” (chap. 8.2)
Press the OK button to enable the setting of the desired function and select the function’s switching state
“1” (function activated) or “0” (function deactivated).
With the Back button you can go back to the “Setpoints” submenu and, repeatedly pressing it, to
Normal display.
In Normal display the activated functions are indicated by an icon showing the number of the respective
function. If several functions are simultaneously active, the icon displays a combination of the
corresponding numbers, e.g.