VD (E3.1) 10/2020 page 36/179
Menu Required access level Functions
Service “Service”
Configuration settings (only for BINDER Service)
Password changing for User and Admin
USB Export: Any user
Import: “Admin”
Export of configuration, logger, and service data
Import of configuration data
Unless noted otherwise, the figure in this manual show the functional range, which is available for the
user with “Admin” authorization level
Note: When specifying the path to the respective function, the possibly required entry of a password is not
listed.
5.2 MB2 controller
The MB2 chamber controller controls following parameters inside the chamber:
Temperature in °C or °F
Pressure in mbar
You can enter the desired set point values in fixed value operation mode directly on the display surface or
via the setpoint menu. For program operation the controller offers programming week and time programs.
In addition, there is a timer program available (stopwatch function).
The controller offers various notifications and alarm messages with visual and audible indication and
remote alarms via e-mail, an event list (trace file) and the graphical display of the measuring values in the
chart recorder view. The MB2 program controller permits programming temperature and pressure cycles,
and specifying special controller functions for each program section. You can enter values or programs
directly at the controller or use the APT-COM™ 4 Multi Management Software (option) specially
developed by BINDER.
Operating mode
Temperature values
Pressure values
Actual values
Set-point values
Functional icons
Figure 17: Normal display of the MB2 program controller (sample values)