VD (E3.1) 10/2020 page 52/179
7.1.2 Pressure set-point entry
Required access level: “User”.
Path: Normal display Setpoints Pressure
Press the OK button to enable the setting.
Pressure setting
The current setting flashes. Enter the desired set-point with the arrow
buttons.
Confirm the entry with the OK button.
Pressure (mbar)
With the arrow-up button you can go back to the temperature set-point entry (chap. 7.1.1).
With the arrow-down button you can now change to the special controller functions setting (chap. 8).
With the Back button you can go back to the “Setpoints” submenu and, repeatedly pressing it, to
Normal display.
7.2 MB2 controller
In Fixed value operating mode, you can enter a temperature set-point, a pressure set-point, and the
switching-state of special controller functions.
All settings made in Fixed value operating mode remain valid until the next manual change. They are
saved also when turning off the chamber or in case of toggling to Program Mode.
The values entered in Fixed value mode become valid again after the end of a program and
will then be targeted and equilibrated.
When operating without a vacuum by setting “Pressure control off” (chap. 8.6.2), the pressure
tolerance range function is automatically deactivated.
7.2.1 Set-point entry through the “Setpoints” menu
Press the Setpoint setting icon to access the “Setpoint” setting menu from Normal display.