Binder KBF-S 720 Humidity Test Chamber

User Manual - Page 59

For KBF-S 720.

PDF File Manual, 91 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
KBF-S / KBF-S-UL (E6) 02/2019 page 59/91
Heat emission of electrical devices connected inside the chamber may modify the temperature
and humidity range.
If the set points for temperature or humidity are outside the optimum range, condensation can arise in the
door area.
CAUTION
Condensation by excess humidity.
Danger of corrosion on the housing after operating at humidity values > 70 % r.h. for
a long period.
Dry the appliance completely before shut-down:
Set the humidity to 0 % r.h. The humidity system must be activated.
Set the temperature set point to 60 °C / 140 °F for approx. 2 hours.
Only then, shut down the chamber at the main power switch (1)
and empty the
freshwater can.
Having turned off the chamber by the main power switch (1), empty the freshwater can.
If you operate the chamber at high humidity and then immediately turn off the chamber, the condensate
collection pan may overflow due to the condensate. This may lead to the emergence of water at the
chamber.
Regularly check the filling level of the condensate collection pan.
18.1 Function of the humidifying and dehumidifying system
Humidifying system
The humidifying and dehumidifying system is located in the humidity generation module. In a cylindrical
container with a volume of approx. 0,07 liters an electrical resistance heating evaporates water. The water
content is kept close to the boiling point, and thus steam can be immediately generated in sufficient quan-
tity for rapid humidity increases or for compensation of humidity losses, e.g. by door openings. Condensa-
tion forming on the outer walls of the useable volume is led through a water drain in the outer chamber to
the wastewater pipe.
Freshwater
Freshwater is supplied by manually filling a freshwater can. The freshwater can be placed on top of the
chamber or in an optionally available magnetic support.
In order to ensure accurate humidifying, observe the following points with regard to the
freshwater supply:
Water type: deionized (demineralized) water
To ensure humidification during 24 hours even at high humidity set-points, we recomm
end
filling the freshwater can at the end of each day.
Water intake temperature NOT below +5 °C / 41 °F and not exceeding 40 °C / 104 °F.
BINDER GmbH is NOT responsible for the water quality provided by the customer.
Any problems and malfunctions that might arise following use of water of deviating quality is
excluded from liability by BINDER GmbH.
Loading ...
Loading ...
Loading ...