Binder KBF-S 720 Humidity Test Chamber

User Manual - Page 25

For KBF-S 720.

PDF File Manual, 91 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
KBF-S / KBF-S-UL (E6) 02/2019 page 25/91
The freshwater can outlet must NOT be placed on or above the rear panel of the chamber.
DANGER
Entry of water via the rear panel into the chamber
Electrical hazard.
Danger of death.
Damage to the chamber.
The regular support may only be attached on the left side of the chamber.
The magnetic support adjustable in height (option) may only be attached on the front or
on the left side of the chamber.
Do NOT place the freshwater can tap on or above the rear panel of the chamber
To operate the chamber with humidity, the freshwater can tap must be open
(chap. 6).
Do not firmly close the lid of the freshwater can to allow air to enter the freshwater
can. As a result, obstruction of the water supply due to negative pressure inside the
can is avoided.
Figure 13: Freshwater can with open tap