Binder KBF-S 720 Humidity Test Chamber

User Manual - Page 52

For KBF-S 720.

PDF File Manual, 91 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
KBF-S / KBF-S-UL (E6) 02/2019 page 52/91
15. Data recorder
An internal data recorder saves chamber data and events in three data sets.
With the export function “Export recorder data” (chap. 16.3) you can save the three data sets via the USB
interface to USB stick. They are issued in the selected language as a spreadsheet with the file extension
“.csv” and can be further processed in the desired program. The data is unencrypted. Always the entire
data memory is read out.
15.1 Recorded data
All data is given out in tabular form. The headlines of the values “number”, “date”, and “time” are given out
in the selected language, all other information in English.
Chamber data for the user “DL1”
Tabular representation of the actual temperature and humidity values together with the date and time,
according to the set storage rate (chap. 15.3). Temperature values are always given out in °C.
Chamber data for BINDER Service “DL2”
This data is intended for use by BINDER Service. It also contains information from the self-test func-
tion. The storage rate is fix (1 minute). Temperature values are always given out in °C.
Event list
Messages regarding the controller and data memory as well as the alarm messages together with the
date and time:
Firmware update done
“New config (USB)”: New configuration uploaded via USB
Data recorder cleared “ Data recorder and event list deleted via setup program
Other event messages according to existing alarms
The moment of switching the alarm state on and off is indicated under “On/Off”.
15.2 Storage capacity
The storage capacity of the data recorder depends on the number of entries.
DL1 = 110.000 entries (equaling 76 days with a storage rate of 1 minute, setting see chap. 15.3)
DL2 = 27.000 entries (equaling 18 days with a fix storage rate of 1 minute)
Event list: 200 events
The shorter the set storage rate, the closer are the stored measuring points, the more precise, but also
shorter is the documented period.
Once the storage capacity of the data recorder is reached, overwriting of the oldest values begins
Loading ...
Loading ...