13.1.1 Connections and forwarding of notifications and alarms to customer systems
The following signals are available to the customer for further processing:
Terminal
Signal
1-2 Alarm threshold 1 O
2
sensor (binary output)
3-4 bridged
Analog signal for O
2
sensor (4 - 20 mA) 0 vol.-% - 35 vol.-% (detection limit 2 vol.-
%)
5-6 Alarm threshold 1 H
2
sensor (binary output)
7-8 bridged Analog signal for H
2
sensor (4 - 20 mA) 0 ppm -1000 ppm (detection limit 5 ppm)
9-10 Alarm threshold 1 CO sensor (binary output)
11-12 bridged Analog signal for CO sensor (4 - 20 mA) 0 ppm - 300 ppm (detection limit 3 ppm)
13-14 Fault notification (binary output), e.g., wire break, sensor defect, flow meter, etc.
15-16 Indicator light green (binary output): Normal operation
17-18 Indicator light yellow (binary output): Alarm threshold 1 triggered, or fault notifica-
tion
19-20 Indicator light red (binary output): Alarm threshold 2 triggered, fire suppression de-
vice activated
21-22 Inertization solenoid valve (binary input): Connection of customer-supplied control
system for flushing/inertization control
Figure 23: Terminal assignment
The bridges of the terminals may only be removed if the respective analog signal shall be
used for transmission.