Binder LIT MK 720 MK 720-liter dynamic climate chambers for rapid temperature changes

User Manual - Page 135

For LIT MK 720.

PDF File Manual, 173 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
MK LIT (E5) 01/2021 page 135/172
22.2 Setting the parameters
This menu allows setting the storage interval, the type of values to be shown and the scaling of the tem-
perature charts.
Path: Main menu > Settings > Measurement chart
“Measurement chart” submenu.
Select the field Storage intervaland enter the desired storage interval. Confirm entry with Confirm
icon.
The available presentation depends on the pre-selected storage rate. Factory setting: 60 seconds. This
means the higher the storage rate, the more precisely but shorter the data representation will be.
In the field Storage valuesselect the desired
value type to be displayed.
For scaling the representation select the desired minimum and maximum temperature value and enter
the desired value. Temperature display range: -50 °C (MK) up to 110 °C. Confirm entry with Confirm
icon.
Setting the storage rate or rescaling (minimum and/or maximum) will clear the measured-value memory
and the event list.
NOTICE
Danger of information loss when setting the storage rate or rescaling.
Data loss of measured-value memory and event list.
Change the storage rate or scaling ONLY if the previously registered data is no longer
needed.
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press
the Close icon to exit the menu without taking over the entries.