Binder LIT MK 720 MK 720-liter dynamic climate chambers for rapid temperature changes

User Manual - Page 14

For LIT MK 720.

PDF File Manual, 173 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
MK LIT (E5) 01/2021 page 14/172
Damage to lithium-ion batteries can lead to an atmosphere inside the battery test chamber that is harmful
to health.
DANGER
Risk of occurrence of toxic gases due to damaged or burning cells/modules/sys-
tems. Risk of self-ignition of cells, modules, or battery systems with flue gas (CO)
development. Risk of emission of hazardous gases from the interior of the cham-
ber.
Death by suffocation. Poisoning.
Shut the entire test down when the first signs of a EUCAR hazard level 4 to 6 event
are detected.
Ensure ventilation measures: Active extraction (technical ventilation in accordance
with national regulations) must be connected to the exhaust port on the chamber. Ex-
traction must remain effective throughout operation of the chamber.
Observe the relevant regulations for handling lithium-ion batteries.
During and shortly after operation, the temperature of the inner surfaces almost equals the set-point. The
window, the access ports and the inner chamber will become hot during operation.
CAUTION
Danger of burning by touching hot chamber parts during operation.
Burns.
Do NOT touch the inner surfaces, the front panel around the inner chamber, the win-
dow, the access port area or the charging material during operation.
1.7.3 Notes on handling CO
2
The CO
2
purge when the CO
2
fire suppression device is activated can create a harmful atmosphere inside
the chamber.
Carbon dioxide (CO
2
) in high concentrations is hazardous to health. It is colorless and almost odorless and
therefore practically imperceptible. An active extraction system (technical ventilation according to country-
specific regulations) must be connected to the exhaust air nozzle of the appliance. The extraction must be
effective during the entire operation of the chamber.
Vent out any CO
2
gas that may escape via good room ventilation or a suitable connection to an exhaust
system. We recommend installing a CO
2
warning system.
DANGER
Danger of suffocation and poisoning by high concentration of CO
2
(> 4 Vol.-%).
Death by suffocation.
Do NOT set up the chamber in non-ventilated recesses.
Ensure technical ventilation measures: An active extraction system (technical ventila-
tion in accordance with country-specific regulations) must be connected to the exhaust
air connection of the chamber. The extraction must be effective during the entire oper-
ation of the chamber.
Observe the relevant regulations for handling CO
2
.
Loading ...