Binder LIT MK 720 MK 720-liter dynamic climate chambers for rapid temperature changes

User Manual - Page 73

For LIT MK 720.

PDF File Manual, 173 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
MK LIT (E5) 01/2021 page 73/172
12.2 Direct setpoint entry via Normal display
Alternatively you can also enter the setpoints directly via Normal display.
Normal display.
Select the setpoint you want to change.
Example: “Temperature” entry menu.
Enter the desired setpoint and confirm entry
with Confirm icon
12.3 Special controller functions via operation lines
Press the Setpoint setting icon to access the “Setpoint” setting menu from Normal display.
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
Operation line Idle modeactivates / deactivates the operating mode Idle mode”.
Operation lines Switching output 1” up to “4” can be used to turn on and off any equipment connected
to the zero-voltage switching outputs (DIN sockets (17) and (18) (option, chap. 25).
Operation line “Compressed air valve” serves to open the solenoid valve of the compressed air connec-
tion (for optional compressed air connection, chap. 26.5)
Operation line Anti-condensationserves to switch the anti-condensation protection (chap. 24).
The other operation lines are without function.
Loading ...
Loading ...
Loading ...