Binder BF 720 Standard Incubator

User Manual - Page 39

For BF 720. Also, The document are for others Binder models: BD 56, BD 115, BD 260, BD 720, BF 56, BF 115, BF 260, ED 56, ED 115, ED 260, ED 720, FD 56, FD 115, FD 260, FD 720, FED 56, FED 115, FED 260, FED 720

PDF File Manual, 107 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
BD / BF / ED / FD / FED (E3.1) 03/2019 page 39/106
The minute value flashes.
Enter the minutes of the timer run-time (0 up to 59)
and confirm with OK.
Chambers without fan (BD, ED):
The timer functionDelayed Offis activated
Back to Normal display.
Chamber with fixed fan speed (FD):
Select whether the fan shall operate or not after the timer has expired
The current setting of the fan operation is shown
Select fan operation: On (100 %) or Off (0 %)
and confirm with OK.
The timer function “Delayed Off” is activated
Back to Normal display.