BD / BF / ED / FD / FED (E3.1) 03/2019 page 15/106
3.2 Guidelines for safe lifting and transportation
The front castors of chambers size 720 can be blocked by brakes. Please move the chambers with cas-
tors only when empty and on an even surface, otherwise the castors may be damaged. After operation
please observe the guidelines for temporarily decommissioning the chamber (chap. 14.2).
CAUTION
Sliding or tilting of the chamber.
Damage to the chamber.
Risk of injury by lifting heavy loads.
Transport the chamber only in its original packaging.
Secure the chamber with transport straps for transport.
Do NOT lift or transport the chamber using the door handle or the door.
Do NOT lift chambers size 720 by hand
Lift chamber size 56 and 115 at its four lower corners with the aid of 2 people,
chamber size 260 with the aid of 4 people, and place it on a transport pallet with
wheels. Push the pallet to the desired site and then lift the chamber from the pallet at
its four lower corners.
Place chamber size 720 using technical devices (fork lifter) on the transport pallet.
Set the fork lifter only from the rear in the middle of the chamber. Make sure to place
all the lateral supports of the chamber on the forks.
Transport chamber size 720 ONLY with the original transport pallet. Set the fork lifter
only to the pallet. Without the pallet the chamber is in imminent danger of overturn-
ing!
Permissible ambient temperature range during transport: -10 °C to +60 °C / 14 °F to 140 °F.
You can order transport packing and pallets for transportation purposes from BINDER Service.
3.3 Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 14.2).
Permissible ambient temperature range during storage: -10 °C to +60 °C / 14 °F to 140 °F.
Permissible ambient humidity: max. 70 % r.H., non-condensing
When after storage in a cold location you transfer the chamber to its warmer installation site, condensa-
tion may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature
and is completely dry.
3.4 Location of installation and ambient conditions
Set up the chamber on an even and non-flammable surface, free from vibration and in a well-ventilated,
dry location and align it using a spirit level. The site of installation must be capable of supporting the
chamber’s weight (see technical data, chap. 16.4 to 16.7). The chambers are designed for setting up
inside a building (indoor use).
CAUTION
Danger of overheating.
Damage to the chamber.
Do NOT set up chambers in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.