M (E2) 04/2019 page 55/71
Chamber size
53
115
240
400
720
Temperature data
Temperature range, 5 °C above ambi-
ent up to
°C
°F
300
572
300
572
300
572
300
572
300
572
Temperature fluctuation
± K
0.3
0.3
0.3
0.3
0.3
Temperature uniformity
(variation)
at 70 °C
± K
0.5
0.6
0.8
0.7
0.7
at 150 °C
± K
1.3
1.5
1.5
1.5
1.9
at 300 °C
± K
2.8
2.8
2.8
5
4.6
Heating up time
to 70 °C
min
5
5
6
6
7
to 150 °C
min
15
16
19
18
21
to 250 °C
min
35
36
42
44
51
Recovery time after door
was opened for 30 sec 2)
to 70 °C
min
1
1
1
1
1
to 150 °C
min
3
3
3
3
3
to 300 °C
min
5
5
5
5
5
Air change data
Air change (air flap open)
at 70 °C
x/h
180
87
57
51
33
at 150 °C
x/h
192
96
60
54
36
at 300 °C
x/h
160
78
54
48
29
Electrical data
IP system of protection acc. to EN 60529
20
20
20
20
20
Nominal voltage (±10 %) 50/60 Hz
V
230 1N~
230 1N~
230 1N~
400 3N~
400 3N~
Nominal power
kW
1.20
1.60
2.70
3.40
5.00
Power plug
shock
proof plug
shock
proof plug
shock
proof plug
CEE plug
5 poles
CEE plug
5 poles
Chamber fuse 5x20 mm / 250V / time-
lag T
Amp
10
external
10
external
16
external
--- ---
Over-current release category B, 3
poles
Amp --- --- ---
16
internal
16
internal
Installation category acc. to IEC 61010-1
II
II
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
2
2
Environment-specific data
Energy consumption
at 70 °C
Wh/h
145
230
370
520
570
at 150 °C
Wh/h
300
544
850
1200
1320
at 300 °C
Wh/h
720
1100
1400
2340
2600
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +25 °C / 77 °F and a power supply voltage fluctuation of ±10. Technical data is determined in accord-
ance to BINDER Factory Standard Part 1:2015 following DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right
to change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.
Loading ...
Loading ...
Loading ...