Binder M 400

User Manual - Page 28

For M 400.

PDF File Manual, 71 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
M (E2) 04/2019 page 28/71
You can enter Program sections into this program table.
Hit the PGM button. An inquiry display appears allowing you to enter or delete individual program sec-
tions:
ZP-Abschnitt Abs. Nr. 5
insert
delete
In this view, new program lines can be entered or deleted:
new New lines are added below in the table
insert New lines are added above a previously selected line
delete Individual lines that have been selected previously are deleted
Create as many lines, i.e. program sections, as desired. As a next step, values can be entered into these
lines. It is possible to add supplementary lines later or to delete individual lines at any time.
To enter values, select the corresponding line via the
arrow keys.
Hit the ENTER button. The program editor ap-
pears.
Enter the individual values of the selected program section.
Program editor Abs.Nr. 6
****
.
*
Setpoint 1 +100.0
FAN
Operating contacts
Time 00:45:00
Repeat Section 5
Repeat Number 10
Tol.-band min. -1999.0
Tol.-band max. +9999.0
Parameter set 1
--
Temperature value at the start of the program section
-- Fan speed in %
-- Operating contact (operation line) 1 = air flap open / closed
-- Duration of the program section
-- No. of start section in case of repeat cycles
-- No. of duplicates in case of repeat cycles
-- Temperature limits (maximum / minimum temperature). In case
of exceeding: temporary program stop.
-- Pre-selected value (Do NOT change!)
Select the parameters via the arrow keys and confirm by pressing “ENTER”..
Then enter the values via the arrow keys, and confirm the entry by pressing “ENTER”..
Adapt the temperature safety device class 2 (chap. 10.1) or the temperature safety device
class 3.1 (option, chap. 10.2) to the highest temperature set-point value of the program actual-
ly used. Check the safety device for each temperature program and adapt it if necessary.
Set the set-point of temperature safety device class 2 or class 3.1 (option) by about 5 °C to 10
°C above the controller temperature set-point.
Performance after completing the program:
The controller changes to Idle Mode. The heating is inactive; the chamber approximates ambient temper-
ature. The fan turns at a 50% rate.
PGM
08:43:55 15.02.02
PGM
08:43:55 15.12.13