11.7 Additional measuring channel for digital object temperature indicator with
flexible temperature sensor Pt 100 (option)
The object temperature display enables the determination of the actual temperature of the charging mate-
rial during the whole process. The object temperature is measured via a flexible Pt100 temperature sen-
sor and can be viewed at the display controller MB1. The sensor top protective tube of the flexible Pt 100
can be immersed into liquid substances.
W
X
CONFIG
HAND
PGM
VIEW->
09:12:24 15.12.13
TEMP
0.0
25.3
°C
OBJ-T
25.6
°C
Figure 13: Display controller MB1 with object temperature display
The object temperature data are put out together with the data of the temperature controller to the RS
422 interface as second measuring channel and can be documented by the APT-COM™ 4 Multi Mana-
gement Software (option, chap. 11.1) developed by BINDER.
Technical data of the Pt 100 sensor:
Three-wire technique
Class B (DIN EN 60751)
Temperature range up to 320 °C / 608 °F
Stainless steel protective tube 45 mm length, material no. 1.4501
11.8 Mostly gas-tight version (option for M 53 and M 115)
With this option the chamber is additionally sealed, so the loss when introducing gases is decreased. The
chamber is not completely gas-tight, so it is impossible to establish overpressure. The sealing diminishes
the release of vapors via the housing that may be set free from the charging material when heated. Carry-
ing-off via the regular evacuation duct, e.g. into a waste air installation, is likely to further reduce emis-
sions.
The chamber is not completely gas-tight. Gases from inside the chamber can escape into the
surrounding atmosphere.
Observe the occupational exposure limit OEL for the released substance set by the national
authorities (formerly maximum permitted workplace concentration). Respect the relevant
regulations.
Any harmful gas that might escape has to be led out via good room ventilation or a suitable
exhaust system. Place the chamber, if necessary, below a gas vent.
The air flap does not close the exhaust duct completely. The delivered plug serves to avoid emerging of
vapors or loss of introduced inert gas, if any, via the exhaust duct. Due to special demands of heat re-
sistance, use the delivered plug only.