Error Messages
Cause And Solution
CAZON
UPPER ARM TYPE
BLOOD PRESSURE MONITOR

1
Error Display
Solution
Sensor abnormality
When measuring, avoid the environment.
(e.g. to magnetic fields, electromagnetic
fields, external electrical influences etc.).
Err 1
Cause

2
Error Display
Cause
Solution
1. No pulse or blood pressure reading due to
very weak artery pulse.
2. The cuff is too loose.
3. Clothing is interfering with the cuff.
1. Remove cuff. Wait 2-3 min and take another
measurement. Repeat steps listed under "BLOOD
PRESSURE MEASUREMENT METHOD"
Iflate the cuff 40mmHg higher than the previous
inflation value.
2. Apply the cuff correctly,Refer to "Wearing cuff
3. Remove any closing interfering with the cuff.
Refer to "Wearing cuff''.
Err 2

4
Error Display
Err 4
Cause
1. Air plug disconnected
2. Air cuff not applied correctly
3. Air is leaking from the cuff
4. Others
Solution
1. Insert the plug securely, Refer to "Wearing cuff".
2. Apply the cuff correctly. Refer to "Wearing cuff".
3. Replace the cuff with the new one.
4. Contact the Customer Service.

5
Error Display
Err 5
Cause
Solution
1. Pipe bend causing air blockage
2. Connector blockage
3. Others
1. Apply the cuff correctly. Refer to "Wearing cuff"
2. Replace the connector
3. Contact the Customer Service Team

6
Error Display
Solution
Cause
Err 6
You can blow into this port
OR use your hairdryers to
clean the dust.
There may be some dust on
the cuff connector port.

7
Error Display
Err 7
Solution
Cause
The cuff was inflated above
300 mmHg when inflating the
cuff manually.
Do not inflate the cuff above 300mmHg.
Refer to "BLOOD PRESSURE
MEASUREMENTMETHOD".