
Display

| Symbol | Function/Meaning | Recommended Action |
|---|---|---|
![]() |
Appears while measurement is in progress. It blinks when the pulse is detected. | Measurement is in progress. Remain as still as possible. |
![]() |
I.H.B symbol appears when an irregular heartbeat is detected. It may appear when a very slight vibration is visible. | |
![]() |
Appears when a body or arm movement is detected. | |
![]() |
Previous measurement is stored in memory. | |
![]() |
Average data | |
![]() |
FULL BATTERY The battery power indicator during measurement. |
|
![]() |
LOW BATTERY The battery power is low when it blinks. |
Replace all batteries with new ones when the symbol blinks. |
| AM | Time in the morning | |
| PM | Time in the afternoon | |
| EE | Device internal error | Remove the Batteries and press the START/STOP button, and then install the batteries again. If the error still appears, contact the customer service. |
| E9 | Unstable blood pressure due to movement during measurement. | Take another measurement. Remain still during measurement. |
| E1 or E2 | The systolic and diastolic values are within 10 mmHg of each other. | Apply the cuff correctly, and take another measurement. |
| The pressure value did not increase during the inflation. | ||
| E3 | The cuff is not applied correctly. | |
| E | PUL DISPLAY ERROR The pulse is not detected correctly |
| Symbols | Function/Meaning |
![]() |
Standby and Turn the device on |
| SYS | Systolic blood pressure in mmHg |
| DIA | Diastolic blood pressure in mmHg |
| PUL/min | Pulse per minute |
![]() |
Battery installation guide |
![]() |
Direct current |
![]() |
Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks |
| SN | Serial number |
![]() |
Refer to instruction manual/booklet |
| IP | International protection symbol |
![]() |
Keep dry |
![]() |
Clock Setting |
![]() |
Clock adjustment and memory recall |

blinks on the display, replace all batteries with new ones. Do not mix old and new batteries. It may shorten the battery life, or cause the device to malfunction. Replace the batteries two seconds or more after the device turns off.
does not appear when the batteries are drained.Insert the air connector plug into the air socket firmly.

| Symbol | Function/Meaning | Recommended Action |
|---|---|---|
![]() |
Artery Position Symbol | Set the O symbol on the artery of the upper arm or in line with the ring finger on the inside of the arm. |
![]() |
Index | |
![]() |
Catalog number | |
![]() |
Lot number |
Note: During measurement, it is normal for the cuff to feel very tight. (Do not be alarmed).


blinks, indicating that the measurement is in progress. Once the pulse is detected, the symbol blinks with each pulse.If re-inflation occurs repeatedly, use the following methods. If your systolic pressure is expected to exceed 230 mmHg, use this procedure.

The device automatically stores up to sixty blood pressure and pulse measurements in memory. Data stored in memory are assigned a data number in the order of the newest to the oldest. The oldest data displays as "No. 1". The
symbol in the upper left corner of the display indicates that you are viewing previous data stored in memory.

Note: If this device stores the two measurements, the average of the two measurements is displayed. If this device stores only one measurement, this is displayed as the average.
symbol blinks.Set the date and time prior to use.

button, the year starts blinking.
button. Press the ◀ button to set the current year and move to month/day selection. The date can be set anywhere between the years 2018 and 2036.
button to set the current month and move to day selection.
button to set the current day and move to hour/minute selection.
button to set the current hour and move to minute selection.
button to turn the device off.Pressing the START/STOP button will turn the device off anytime.
Note: After three minutes of non-operation, the device will turn off automatically.
When the clock has not been set, the clock display indicates dashes.
When using the device for the first time, the clock is not adjusted. When the device is disconnected from the power supply, the set date and time will be erased. When the set date and time is erased, please adjust again.
The blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 15% from the average of all heartbeats during the blood pressure measurement. It is important that you are relaxed, remain still and do not talk during measurements.
Note: We recommend contacting your physician if you see this
indicator frequently.
The indicator monitors the progress of pressure during measurement.

What is Blood Pressure?
Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg). One's natural blood pressure is represented by the fundamental pressure, which is measured first thing in the morning while one is still at rest and before eating.
Each segment of the bar indicator corresponds to the AHA blood pressure classification.
AHA Classification Indicator

: The indicator displays a segment, based on the current data, corresponding to the AHA classification.
| Problem | Possible Reason | Recommendation Action |
|---|---|---|
| Nothing appears on the display, even when the power is turned on. | Batteries are drained. | Replace all batteries with new ones. |
| Battery terminals are not in the correct position. | Reinstall the batteries with negative and positive terminals matching those indicated on the battery compartment. | |
| The cuff does not inflate. | Battery voltage is too low. blinks. If the batteries are drained and completely, the symbol does not appear. |
Replace all batteries with new ones. |
| The device does not measure. Readings are too high or too low. |
The cuff is not applied properly. | Apply the cuff correctly. |
| You moved your arm or body during measurement. | Make sure you remain still and quiet during measurement. | |
| The cuff position is not correct. | Sit comfortably and still. Place your arm on a table with your palm facing upward and the cuff at the same level as your heart. | |
| If you have a very weak or irregular heartbeat, the device may have difficulty in determining your blood pressure. | ||
| Other | The valve is different from that measured at a clinic or doctor's office. | At a clinic or doctor's office, apprehension may cause an elevated reading. Home measurement reduces the effects of outside influences on blood pressure readings & supplements the doctor's readings. |
| Remove the batteries. Place them back properly and take another measurement. |
Note: If the actions described above do not solve the problem, contact customer service. Do not attempt to open or repair this product, as any attempt to do so will make your warranty invalid.
Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, contact customer service.
| Type | UA-4000WM |
|---|---|
| Measurement method | Oscillometric measurement |
| Measurement range | Pressure: 0 - 299 mmHg Systolic pressure: 60 - 279 mmHg Diastolic pressure: 40 - 200 mmHg Pulse: 40 - 180 beats/minute |
| Measurement accuracy | Pressure: ±3 mmHg Pulse: ±5 % |
| Power Supply | 4 x 1.5V batteries (R6P, LR6 or AA) |
| Number of measurements | Approx. 700 LR6 or AA (alkaline batteries) Approx. 200 R6P (manganese batteries) With pressure value 180 mmHg, room temperature 23 °C. |
| Classification | Internally powered ME equipment (by batteries) Continuous operation mode |
| Clinical test | According to ISO81060-2-2013 |
| EMC | IEC 60601-1-2: 2014 |
| Memory | Last 60 measurements |
| Operating conditions | +10 to +40 °C/15 to 85 %RH/800 to 1060 hPa |
| Transport/Storage conditions | -20 to +60 °C/10 to 95 %RH/700 to 1000 hPa |
| Cuff Circumference | UA-420W/M: 8.6-16.5" (22 to 42 cm) |
| Dimensions | Approx. 3.94"W"×2.68"H"×5.47"D" (110"W"×68"H"×138"D") |
| Weight | Approx. 9.7 oz/275 g, excluding the batteries |
| Ingress protection | Device: IP20 |
| Applied part | Cuff: Type BF |
| Useful life | Device: 5 years (when used six times a day) Cuff: 2 years (when used six times a day) |
Note: Specifications are subject to change without prior notice. IP classification is the degree of protection provided by enclosures in accordance with IEC 60529. This device is protected against solid foreign objects of 12 mm diameter and greater such as a finger. This device is not protected against water.
| Product | Warranty Term |
| Monitor | 5 year |
| Cuff | 2 year |
Limited Warranty:
This warranty warrants to the first purchaser ("You") that the product You purchased (the "Product") will be free from defects in material, workmanship and design for the applicable Warranty Term stated above from the date You purchased the Product under normal use. This Limited Warranty is personal to You and is not transferable. If the Product is defective, then You return the Product in accordance with the procedure set forth below. This warranty obligation is limited to the repair or replacement, at our option, of the defective Product that has been returned by You within the warranty period. Such repair or replacement will be at no charge to You. The repaired or replacement Product is warranted here-under for the longer of the remainder of the original warranty period or 90 days from the date of shipment of the repaired or replacement Product.
To obtain a warranty service, please contact us at 1-888-726-9966 for return address, shipping and handling fee, and other instructions for processing warranty. Please ensure you have satisfactory proof of the date of Your purchase and a description of the defect. Returns will not be accepted unless a Return Material Authorization (RMA) Number has been issued from our Customer Service Representative.
This Limited Warranty does not cover, and we will not be liable for (i) any shipment damage, (ii) any damage or defect due to misuse, abuse, failure to use reasonable care, failure to follow written instructions enclosed with the Product, accident, subjecting the Product to any voltage other than the specified voltage, improper environmental conditions, or modification, alteration or repair by anyone other than persons authorized, or (iii) expendable or consumable components.
THIS LIMITED WARRANTY IS THE ONLY WARRANTY PROVIDED; THERE ARE NO OTHER EXPRESS WARRANTIES.
The above remedy of repair or replacement is your only and exclusive remedy. IN NO EVENT SHALL WE BE LIABLE FOR ANY DAMAGES, INCLUDING, WITHOUT LIMITATION, LOST PROFITS, LOST INFORMATION OR REPLACEMENT COSTS, ARISING OUT OF YOUR USE OF OR INABILITY TO USE THE PRODUCT, INCLUDING, WITHOUT LIMITATION, ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF WE HAVE BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
Some states do not allow the exclusion of incidental or consequential damages, so that the above exclusions may not apply to you. This warranty gives you specific legal rights, and you may also have other rights that may vary from state to state.
No distributor, dealer or other party is authorized to make any warranty or to modify this warranty, or to assume any liability with respect to its products.
Satisfaction guaranteed — Or we'll replace it or give you your money back. For questions or comments or to report an undesired reaction or side effect, please call 1-888-287-1915.
Walmart Inc.
702 S.W. 8th St.
Bentonville, AR 72716
www.walmart.com
1-888-287-1915
Reference file: Equate UA-4000WM Upper Arm Blood Pressure Monitor and Wide Range Arm Cuff
Series: 4000 SERIES
Additionally, the document applies to other Equate models: VIPRB-UA-4000WM, 1WMPD4004758A