Rigol DS1102E oscilloscope

Rigol DS1000E/D user manual - Page 81

For DS1102E.

PDF File Manual, 172 pages, Read Online | Download pdf file

DS1102E photo
Loading ...
Loading ...
Loading ...
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-51
Pattern Trigger (DS1000D Series)
Pattern trigger identifies trigger terms by checking appointed code. The code is
logical relationship of all channels, with high (H), low (L) and ignore(X).
Figure 2- 59 Table 2- 35 The Pattern Trigger menu
Menu
Settings
Comments
Select
D15-D0
Choose digital channel for Pattern trigger
Code
H
L
X
High
Low
Ignore
Rising Edge
Falling Edge
Sweep
Auto
Normal
Single
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform
and then stop
Set Up
To go to set up menu, see table 2-40
Key Points:
H (High): Logic high: voltage is higher than threshold setting.
L (Low): Logic low: voltage is lower than threshold setting.
X (Ignore): Dont care. If all the channels are ignored, the oscilloscope won’t be
triggered.
Rising Edge ( ) or Falling Edge ( ): Set the code as an edge of the channel,
rising edge or falling edge. When the edge is appointed, if code settings of other
channels are all true, the oscilloscope will be triggered on the appointed edge. If no
edge is appointed, the oscilloscope will be triggered on the last edge whose code is
true.