Rigol DS1102E oscilloscope

Rigol DS1000E/D user manual - Page 63

For DS1102E.

PDF File Manual, 172 pages, Read Online | Download pdf file

DS1102E photo
Loading ...
Loading ...
Loading ...
RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
2-33
Delayed Scan:
The Delayed Scan is a magnified portion of the main waveform window. Use
Delayed Scan to locate and horizontally expand part of the main waveform window
for a more detailed (higher horizontal resolution) analysis of signal. The Delayed
Scan time base setting cannot be set slower than the Main time base setting.
Figure 2- 39
Delayed Scan window
The following steps show you how to use Delayed Scan.
1. Connect a signal to the oscilloscope and obtain a stable display.
2. Press horizontal MENUDelayedON or press horizontal knob to
enter Delayed Scan mode.
The screen splits into two parts. The upper half displays the main waveform window
and the lower half displays an expanded portion of the main waveform window. This
expanded portion of the main window is called the Delayed Scan window. Two
blocks shaded at the upper half; the un-shaded portion is expanded in the lower half.
The horizontal and knobs control the size and position of
the Delayed Scan. The value at bottom of the screen is the main time base and the
Waveform to be horizontally expanded
Expanded waveform in horizontal
Main timebase
Time base of
Delayed Scan
Loading ...
Loading ...
Loading ...