RIGOL
© 2008 RIGOL Technologies, Inc.
User’s Guide for DS1000E, DS1000D Series
1-7
oscilloscope:
Align the slot in the probe connector with the key on the CH 1 BNC.
Push to connect, and twist to the right to lock the probe in place.
Attach the probe tip and ground lead to the PROBE COMP connector.
Figure 1- 6
Attach the probe
② Set the probe attenuation to 10X. To do this, push CH1→Probe→10X.
Figure 1- 7 Figure 1- 8
Set attenuation on the probe Set attenuation in the menu
③ Push the AUTO button. Within a few seconds, a square wave will display.
④ Push the OFF button or push the CH1 button again to turn off Channel 1.
Push the CH2 button to turn on channel 2, repeat steps 2 and 3.
NOTE: The signal output from Probe compensator should only be used for probe
compensation, not for calibration.
Probe compensator
Probe scale