Chapter XIV Technical Information
Except for those specifications marked with “Typical”, all specifications have warranties.
Unless otherwise specified, all technical specifications are applicable to the probes with
attenuation switch set as 10× as well as UTD2000 series DSO. DSO must first meet the
following two conditions to meet those specification standards:
The instrument must continuously operate for over half an hour within the operating
temperature.
If the change scope of operating temperature is or exceeds 5°C, please perform the
“Self-Adjustment” function in the UTILITY system function.
Basic Specification
Model Analog Bandwidth Rise Time(Typical) Real-time sampling
Equivalence
sampling
UTD2052CL+ 50MHz ≤7ns
500MS/s 25GS/s
UTD2072CL 70MHz ≤5ns
UTD2102CL+ 100MHz ≤3.5ns
UTD2102CL PRO 100MHz ≤3.5ns
UTD2152CL 150MHz ≤2.4ns
UTD2052CEX+ 50MHz ≤7ns
1GS/s 50GS/s
UTD2102CEX+ 100MHz ≤3.5ns
UTD2152CEX 150MHz ≤2.4ns
UTD2202CEX 200MHz ≤1.8ns
Acquire System Specification
Average
When the sampling times of all channels are N, N can be chosen among 2, 4, 8, 16, 32,
64, 128 and 256
Input Channel Specifications
Input Coupling DC, AC and GND
Input impedance (1MΩ± 2%)//(18pF± 3 pF)
Probe attenuation
coefficient
0.01×/0.02×/0.05×/0.1×/0.2×/0.5×/1×/2×/5×/10×/20×/50×/100×/200×/500×/1000×
Maximum input voltage
400Vpk, the transient over voltage is 1000 Vpk.
Horizontal System Specification
Time-base scale 2ns/div-50s/div
Waveform interpolation Sin(x)/x