SDS3000X HD Series Digital Oscilloscope User Manual
68 int.siglent.com
Sequence N-1
Example:
Input a pulse sequence with a period of 50 ms to C1. The rise time of the pulse is 2 ns, while fall time
is 100 ns; pulse width is 108 ns, and amplitude is 1.6 Vpp. Press the AutoSetup button on the front
panel.
Set the coupling mode of C1 to DC50Ω, and vertical scale to 500 mV/div, vertical offset to 0. Set the
trigger level to 0.
In normal mode, 5 pulses can be obtained on the screen with the sample rate of 1GSa/s at
the maximum memory depth.
Set the trigger mode to "Single", the timebase to 100 ns/div. Turn on the Sequence mode, and set the
segments to maximum (74403 in this example, up to 80,000 depending on the number of samples at