16.10 Zone Trigger
The SDS3000X HD includes a zone trigger to help isolate elusive glitches. There are two user-defined
areas: Zone1 and Zone2. Users can set the property of each zone as intersect or not intersect as
an additional condition to further isolate the interesting event quickly. Intersect” only includes events
that occur within the zone. “Not-intersect” events include all events that occur outside of the zone.
Touch Trigger > Zone to recall zone dialog box:
A. Turn on or off zone trigger
B. Select the source: C1~C4
C. Turn on or off zone1
D. Set the property of zone1: Intersect or Not intersect
E. Set the coordinate of zone1. The range is within the
waveform area
F. Turn on or off zone2
G. Set the property of zone2: Intersect or Not intersect
H. Set the coordinate of zone2. The range is within the
waveform area
The zones can be created and moved by gestures or by setting Zone > Zone Setting in the
dialog box. The color of the zones outline is consistent with the color of the specified source (Channel
1 = Yellow, etc...).
Gesture
When the zone trigger is turned on, touch-and-hold on any position within the waveform area and
draw a rectangular box, as shown in the following figure: