SDS3000X HD Series Digital Oscilloscope User Manual
int.siglent.com 109
If you set the trigger condition to 7 address & data or 10
address & data:
Address can be selected in the hexadecimal range of
0x00 to 0x7F (7-bit) or 0x3FF (10-bit). If the address is
selected as "0xXX (7-bit address)" or "0xXXX (10-bit
address)", the address is ignored. It will always trigger
on the Ack bit followed by the address.
Data1 and Data2 can be selected in the
hexadecimal range of 0x00 to 0xFF. If the data is
selected as "0xXX", the data is ignored. It will always
trigger on the Ack bit followed by the address.
R/W Bit can be specified as Write, Read, or Don't
Care.
Data Length -- The data length range is 1 to 12 bytes. When the SDA data length is equal to the
value of Date Length and address length is the same as the set value, the oscilloscope will be
triggered.
Touch the Address Length to select "7-bits" or "10-bits" to match the address of the input
signal.
Touch the Data Length and turn the universal knob or virtual keypad to set the data length to
match the data length of the input signal.