SDS3000X HD Series Digital Oscilloscope User Manual
int.siglent.com 165
19.7 Measurement Statistics
Measurement statistics are based on the total number of captured waveforms. In Roll mode,
measurement statistics increase over time. Touch Statistics Setting in the measure dialog box to
recall the Statistics Config dialog box:
A. Set the maximum number of samples for the statistics function.
The setting range is 1 ~ 1024, or unlimited. If there is no limit, the
number of statistics will be accumulated. If there is a limit, when
the maximum number of statistics N is reached, only the latest N
measurements will be counted. When the count is limited, only the
first period measurement value of each frame is counted
B. Set the upper limit of the statistic number in a frame. Valid only
when the count limit is set to "Unlimited".
C. Clear and restart the statistics. Pressing the Clear Sweeps
button on the front panel has an equivalent effect
D. Turn the histogram on or off
E. Return to the previous menu
Enable the Statistics function to observe the distribution of the measured values of every selected
parameter.
Value -- The current measurement
Mean -- Average of all historical measurements
Min -- The minimum of all historical measurements
Max -- The maximum of all historical measurements
Stdev -- Standard deviation of all historical measurements
Count -- The number of historical measurements
Press the Clear Sweeps button or touch Reset Statistics in the statistics setting dialog box, or
touch the symbol in the statistics display area to clear and restart statistics.