RIGOL Chapter 10 Pass/Fail Test
10-4 MSO2000A/DS2000A User’s Guide
waveforms are detected. The test results on the screen will update
continuously and a pulse will be output from the [Trigger Out] connector
at the rear panel each time a failed waveform is detected.
3. Beeper on Fail
Press Output to select “Fail” or “Fail+ ”.
Fail: when a failed waveform is detected, the beeper does not sound.
Fail+
: when a failed waveform is detected, the beeper sounds (not
related to the on/off state of the sound).
4. [Trigger Out] Outputs Pulse
Press AuxOutput to select “ON” or “OFF”. If “ON” is selected, the oscilloscope
outputs a pulse from the [Trigger Out] connector at the rear panel when a
failed waveform is detected.
You can also press Utility AuxOutput and select “PassFail” to set this
output.
To Save the Test Mask
Users can save the current test mask in “* .pf” format to the internal Flash memory or
external USB storage device. The internal memory can store at most 10 test mask
files (Lo c a l PF. p f ) .
Press Save to enter the file store interface. Please refer to the relative descriptions in
“
Store and Recall” to save the test mask file to the internal or external memory.
To Load the Test Mask
Users can also load the test mask files (* .pf) stored in the internal Flash memory or
external USB storage device to the instrument.
Press Load to enter the file recall interface. Please refer to the relative descriptions
in “
Store and Recall” to load the test masks to the instrument.