Chapter 10 Pass/Fail Test RIGOL
MSO2000A/DS2000A Users Guide 10-3
horizontal and vertical adjustment ranges are 0.02 div to 4.0 div and 0.04 div to 5.12
div respectively.
Figure 10-1 Creat Mask
To Set Test Results Output Type
You can set the output type of the test results referring to the following methods.
1. The Statistic Message
Press Message to select ON or OFF. When ON is selected, the statistic
message of the test results (including the number of the failed frames, the
passed frames and the total frames) will be displayed at the upper right corner
of the screen, as shown in the f igure below.
2. Stop on Fail
Press StopOnFail to select ON or OFF.
ON: when a failed waveform is detected, the oscilloscope will stop the test
and enter the STOP state. At this point, the results of the test remain the
same on the screen (if display is turned on) and only one negative pulse is
output from the [ Trigger Out] connector (if enabled) at the rear panel.
OFF: the oscilloscope will continue with the test even though failed
Loading ...