RIGOL Chapter 5 To Trigger the Oscilloscope
5-32 MSO2000A/DS2000A User’s Guide
Timeout Time:
Press TimeOut to set the timeout time of timeout trigger and the range is from 16
ns to 4 s.
Trigger Mode:
Press Sweep to open the trigger mode list and select “Auto”, “Normal” or ”Single”.
For the details, please refer to “
Trigger Mode”. The corresponding status light of
the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Note: Noise rejection is grayed out and disabled automatically when the signal
source is set to any channel of D0-D15.
Trigger Level:
Use TRI GGER
LEVEL to modify the level. For the details, please refer to the
introduction of “
Trigger Level”.