Rigol MSO2102A-S Oscilloscope

User Manual - Page 118

For MSO2102A-S.

PDF File Manual, 317 pages, Read Online | Download pdf file

MSO2102A-S photo
Loading ...
Loading ...
Loading ...
RIGOL Chapter 5 To Trigger the Oscilloscope
5-32 MSO2000A/DS2000A Users Guide
Timeout Time:
Press TimeOut to set the timeout time of timeout trigger and the range is from 16
ns to 4 s.
Trigger Mode:
Press Sweep to open the trigger mode list and select Auto”, Normal” or Single”.
For the details, please refer to
Trigger Mode. The corresponding status light of
the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Note: Noise rejection is grayed out and disabled automatically when the signal
source is set to any channel of D0-D15.
Trigger Level:
Use TRI GGER
LEVEL to modify the level. For the details, please refer to the
introduction of
Trigger Level”.
Loading ...
Loading ...
Loading ...