ii900/ii910
Mode d’emploi
10
3. If the measured distance requires additional adjustment,
tap the Distance box to update.
4. Press Capture button or Save in the menu.
The Imager saves the still photo of the scene with
overlapping sound image and values in .PNG or .JPG
format. You can add notes, photo notes, or tags to the
image. See Memory for more information.
Note
For best results:
Move the Imager around the leak to find the
highest LeakQ value. This will best reflect the
actual leak size.
Obstacles within the circle may influence the
calculated distance and LeakQ value.
High background noise will influence the calculated
distance and LeakQ value.
PDQ Mode™ (ii910)
PDQ Mode™ captures data from a Partial Discharge (PD) that
enables further analysis such as type of discharge and
severity. The PD must be inside the circle on the display.
When a PD is detected, the PD Count value on the display
gives an indication of the pulses generated by that partial
discharge.
The value is based on the pulses of the acoustic interference
generated by the PD:
Values for estimated DISTANCE show on the display.
or
NO TARGET FOUND shows on the display when no PD
is detected inside the circle.
If the Imager cannot automatically determine the distance, this
pop-up message shows on the display:
UNABLE TO ESTIMATE DISTANCE
If the Imager is unable to determine the distance, or you want
to override the estimated distance, you can input the distance
manually. See LeakQ™ Mode.
Note
PD Count value is not directly related to the distance.
The Imager saves the still photo of the scene with overlapping
sound image and values in .PNG or .JPG format. You can add
notes, photo notes, or tags to the image. See Memory for
more information.
MecQ™ Mode (ii910)
MecQ™ mode detects and locates potential anomalies in
mechanical components as an early identification of possible
mechanical deterioration that requires further inspection.