FLUKE 8845A-EFPT Fine Point Extended Test Probe Adaptor Set

User Manual for the Fluke 8845A-EFPT Fine Point Extended Test Probe Adaptor Set - Page 80

For 8845A-EFPT.

PDF File Manual, 100 pages, Read Online | Download pdf file

8845A-EFPT photo
Loading ...
Loading ...
Loading ...
8808A
Users Manual
4-20
Table 4-10. Function Modifier Commands and Queries (cont.)
Command Description
HOLDTHRESH? Meter returns Touch Hold <threshold> (1, 2, 3, or 4). See “Touch Hold
Function (HOLD)” in Chapter 3 for more information.
MAX Meter enters MAX modifier with present reading as maximum value. If already
in MAX modifier, Meter displays maximum value. In MAX modifier,
autoranging is disabled. See “Minimum / Maximum Modifier (MIN MAX)” in
Chapter 3 for more information.
MAXSET <numeric value> Meter enters MAX modifier with <numeric value> as the maximum value.
<numeric value> can be a signed integer, signed real number without
exponent, or signed real number with exponent. Autoranging is disabled. See
“Minimum / Maximum Modifier (MIN MAX)” in Chapter 3 for more information.
If <numeric value> exceeds the measurement range, an Execution Error is
generated.
MIN Meter enters MIN modifier with present reading as minimum value. If already
in MIN modifier, Meter displays minimum value. In MIN modifier, autoranging
is disabled. See “Minimum / Maximum Modifier (MIN MAX)” in Chapter 3 for
more information.
MINSET <numeric value> Meter enters MIN modifier with <numeric value> as the minimum value.
<numeric value> can be a signed integer, signed real number without
exponent, or signed real number with exponent. Autoranging is disabled. See
“Minimum / Maximum Modifier (MIN MAX)” in Chapter 3 for more information.
If <numeric value> exceeds the measurement range, an Execution Error is
generated.
MNMX Meter enters MIN MAX modifier with present reading as minimum and
maximum value. If already in MIN MAX modifier, Meter displays latest MIN or
MAX value. In MIN MAX modifier, autoranging is disabled. See “Minimum /
Maximum Modifier (MIN MAX)” in Chapter 3 for more information.
When the MIN MAX modifier is selected, you can toggle between displaying
the minimum and maximum readings without losing the minimum and
maximum values stored.
MNMXSET <numeric1,
numeric2>
Meter enters MIN MAX modifier with <numeric1> as the maximum value and
<numeric2> as the minimum value.
<numeric1> and <numeric2> can be signed integer, signed real number
without exponent, or signed real number with exponent. Autoranging is
disabled. See “Minimum / Maximum Modifier (MIN MAX)” in Chapter 3 for
more information.
If <numeric1> or <numeric2> exceeds the measurement range, an Execution
Error is generated.
MMCLR Meter exits the MN MX modifier. The stored minimum and maximum values
are lost, and the Meter returns to the ranging mode and range selected prior
to selecting MN MX modifier.