Operating the Meter Using the Computer Interface
Computer Interface Command Set 4
4-17
Table 4-8. Common Commands (cont.)
Command Name Description
*SRE Service Request Enable Sets the Service Request Enable Register to
<value>, where <value> is an integer between 0 and
255. The value of bit 6 is ignored because the Service
Request Enable Register does not use it.
<value> is an integer whose binary equivalent
corresponds to the state (1 or 0) of bits in the register.
If <value> is not between 0 and 255, an Execution
Error is generated.
*SRE? Service Request Enable
Query
Meter returns the <value> of the Service Request
Enable Register (with bit 6 set to 0).
<value> is an integer whose binary equivalent
corresponds to the state (1 or 0) of bits in the register.
*STB? Read Status Byte Meter returns the <value> of the Status Byte with bit 6
as the Master Summary bit.
<value> is an integer whose binary equivalent
corresponds to the state (1 or 0) of bits in the register.
*TRG Trigger Causes the Meter to trigger a measurement when
parsed.
*TST Self test query Always returns zero.
*WAI Wait-to-Continue Do nothing.
Function Commands and Queries
Table 4-9 describes function commands and queries. Refer to Chapter 3 for detailed
descriptions of each function.
Table 4-9. Function Commands and Queries
Commands
Primary Display Secondary Display
Function
AAC AAC2 AC current
AACDC
[1]
(Not applicable) AC plus DC rms current
ADC ADC2 DC current
(Not applicable) CLR2 Clears measurement (if shown)
CONT (Not applicable) Continuity test
DIODE (Not applicable) Diode test
FREQ FREQ2 Frequency