Specifications
Safety
8
125
Safety
Designed for 1000 V Measurement Category III (Fluke
190-502: 1000V Measurement Category II) , 600 V
Measurement Category IV (Fluke 190-502: 600V
Measurement Category III), Pollution Degree 2, per:
EN/IEC 61010-1:2001 Pollution Degree 2 (According
to CE mark)
IEC61010-031:2002+A1:2008
ANSI/UL 61010-1:2004 {ed. 2.0}
CAN/CSA C22.2 No. 61010-1-04 (including approval)
Max. Input Voltages
FLUKE 190-2xx,
-1xx, -062
FLUKE 190-502
BNC Input A, B,
(C, D) directly
300 V CAT IV 300 V CAT III
Via VPS410 1000 V CAT III
600 V CAT IV
1000 V CAT II
600 V CAT III
Via VPS510 300 V CAT III 300 V CAT III
METER/EXT
banana input
1000 V CAT III
600 V CAT IV
1000 V CAT II
600 V CAT III
Max. Floating Voltage
FLUKE 190-xx4, -102, -202, -062 (test tool or test tool +
VPS410)
From any terminal to earth ground ......... 1000 V CAT III
600 V CAT IV
Between any terminal ............................. 1000 V CAT III
600 V CAT IV
FLUKE 190-502 (test tool or test tool + VPS410)
From any terminal to earth ground .......... 1000 V CAT II
600 V CAT III
Between any terminal .............................. 1000 V CAT II
600 V CAT III
FLUKE 190-xxx + VPS510
From any terminal to earth ground ........... 300 V CAT III
Between any terminal ............................... 300 V CAT III
Voltage ratings are given as “working voltage”. They
should be read as Vac-rms (50-60 Hz) for AC sinewave
applications and as Vdc for DC applications.